Real-Time ECG Artifact Removal for Adaptive Deep Brain Stimulation: A Comparative Study.
Lucrezia Silvi1, Valentina D'Onofrio2, Simone Cauzzo1,2
1Biomedical Engineering Research to Advance and Innovate Translational Neuroscience (BRAIN Unit), Department of Neuroscience, University of Padova, 35122 Padua, Italy.
Template subtraction (TS) is the best method for removing electrocardiographic (ECG) artifacts in real-time adaptive deep brain stimulation (aDBS) for Parkinson's disease (PD). It balances computational efficiency and biomarker accuracy for closed-loop control.
Area of Science:
- Neuroscience
- Biomedical Engineering
- Signal Processing
Background:
- Adaptive deep brain stimulation (aDBS) for Parkinson's disease (PD) requires precise detection of beta oscillatory activity.
- Electrocardiographic (ECG) artifacts commonly contaminate local field potentials (LFPs), hindering the performance of control algorithms.
- The real-time feasibility of existing offline ECG artifact removal methods for aDBS is not well-established.
Purpose of the Study:
- To evaluate the performance of four ECG removal algorithms for real-time aDBS applications.
- To assess artifact removal quality and real-time processing feasibility of different algorithms.
Main Methods:
- Four ECG removal algorithms were tested: template subtraction (TS), singular value decomposition (SVD), extended SVD (eSVD), and Perceive toolbox (PR).
- Simulated datasets with varying contamination levels (-30 to +20 dB) and clinical recordings from 20 PD patients were used.
- Performance metrics included beta power preservation, signal-to-noise ratio, and 99th percentile processing latency (P99 < 50 ms for real-time feasibility).
Main Results:
- Template subtraction (TS) and standard SVD met the real-time feasibility threshold (P99 < 50 ms).
- TS demonstrated superior timing consistency (P99 ≈ 10 ms), while eSVD and PR exceeded the latency threshold (P99 > 90 ms).
- TS maintained beta power accuracy within ±12% across contamination levels, offering the best balance between artifact suppression and signal preservation.
Conclusions:
- Template subtraction (TS) is the recommended algorithm for real-time ECG artifact removal in aDBS for Parkinson's disease.
- TS provides a critical balance of computational efficiency and biomarker fidelity necessary for closed-loop control systems.
- This method ensures reliable detection of beta oscillatory activity essential for effective aDBS therapy.