In Vitro Evaluation of The Effects Of Er,Cr:YSGG and Diode Lasers Used on Titanium Cylinder
Published on: June 6, 2025
External Root Surface Temperature Elevation During Gutta Percha Removal Using Er,Cr:YSGG Laser: A Time-Duration
Annappa Raghavendra Vivekananda Pai1, Ashwini Savia Colaco2
1Department of Conservative Dentistry and Endodontics, Manipal University College Malaysia (MUCM) Faculty of Dentistry, Melaka, Malaysia.
European Endodontic Journal
|May 13, 2026
Summary
Longer erbium, chromium: yttrium, scandium, gallium, garnet (Er,Cr:YSGG) laser application for gutta-percha removal significantly increases root temperature. To prevent tissue damage, Er,Cr:YSGG laser use for gutta-percha removal should not exceed 5-8 minutes.
Area of Science:
- Endodontics
- Dental Lasers
- Root Canal Treatment
Background:
- Root temperature elevation during laser application is a known concern.
- Limited data exists on the relationship between Er,Cr:YSGG laser duration and temperature rise during gutta-percha removal.
Purpose of the Study:
- To comparatively evaluate root temperature changes during gutta-percha (GP) removal using the Er,Cr:YSGG laser at different application durations.
- To determine the optimal duration for safe and effective GP removal with the Er,Cr:YSGG laser.
Main Methods:
- Forty-two gutta-percha-filled root samples were divided into three groups (n=14) for Er,Cr:YSGG laser treatment.
- GP removal was performed using identical laser parameters, with application durations of 8, 10, and 12 minutes.
- External root surface temperature was measured at the apical third using a thermocouple and digital thermometer.
Main Results:
- A significant temperature increase was observed with longer Er,Cr:YSGG laser application durations (P < .05).
- Groups II (10 minutes) and III (12 minutes) exceeded the critical thermal limit for root surface temperature.
- The highest temperature elevation occurred in the 12-minute group, followed by the 10-minute and 8-minute groups.
Conclusions:
- Extended Er,Cr:YSGG laser application for gutta-percha removal leads to temperature elevations exceeding critical thermal limits.
- To minimize detrimental effects on surrounding tissues, Er,Cr:YSGG laser use for GP removal should be limited to 5-8 minutes.
- Further clinical studies are needed to validate these findings and establish best practices for laser-assisted root canal procedures.