Anesthesia Workstation Performance in Pediatric Ventilation: Part 1. Precision of Ventilation
Johannes Spaeth1, Malte Trost1, Bernd Flamm1
1Department of Anesthesiology and Critical Care, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Background:
Mechanical ventilation of children sets the highest demands on the performance of anesthesia workstations. Different ventilator technologies are available, each representing unique pneumatic characteristics. The hypothesis of this study was that precision of delivered ventilation parameters depends on the ventilator type.
Methods:
Six modern anesthesia workstations (GE HealthCare Carestation 650; Mindray Bio-Medical Electronics Co. Ltd. Wato EX-65 Pro and A9; Dräger Medical Atlan A350 and Perseus A500; and Getinge Flow-c) were tested in physical models simulating the respiratory systems of a preterm baby, a neonate, an infant, and a toddler. Volume-controlled and pressure-controlled ventilation were investigated with and without airway leakage. Delivered tidal volume, peak inspiratory pressure, positive end-expiratory pressure, and inspiratory to expiratory ratio were compared to the set values. The data are presented descriptively; a difference greater than or equal to 10% between the delivered and set values was considered clinically relevant.
Results:
Deviations of the delivered volume from the set tidal volume during volume-controlled ventilation ranged between -19.5 ± 0.6% and +35.1 ± 3.0% in the preterm baby model but decreased with increasing tidal volume. Positive end-expiratory pressure and peak inspiratory pressure during pressure-controlled ventilation did not differ relevantly from the set values in any device or lung model. Expiratory ratios varied widely in most devices, leading to relevant differences in tidal volume during pressure-controlled ventilation. With leakage, tidal volume was, on average, 7% lower during volume-controlled ventilation but 2% lower during pressure-controlled ventilation.
Conclusions:
Precision of ventilation with modern anesthesia workstations varies depending on the ventilator type, ventilation mode, and set parameters. Pressure-controlled ventilation is advantageous in the six workstations tested considering the good agreement between the applied and the set parameters and leakage compensation. The results imply precision superiority of pressure-controlled ventilation over volume-controlled ventilation in preterm babies and neonates, independent from the ventilator technology in use.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...