Related Experiment Video
Updated: May 15, 2026

Effects of Surgical Masks on Cardiopulmonary Function in Healthy Subjects
Published on: February 12, 2021
Fit testing of N95 respirators: A pilot study on respirator design and demographic factors
Xinyi Niu1, Michael Yermakov1, Sergey Grinshpun1
1Department of Environmental and Public Health Science, College of Medicine, University of Cincinnati, Cincinnati, Ohio.
Abstract:
N95 filtering facepiece respirators (FFRs) approved by the National Institute for Occupational Safety and Health are widely used by healthcare workers to prevent the inhalation of hazardous airborne contaminants. Various nose-clip shapes and respirator styles in N95 FFRs are designed to eliminate gaps between the respirator and the wearer's facial skin, ensuring adequate protection for the wearer. However, variations in facial dimensions across sexes and races may affect the fit and, consequently, the performance of these respirators. The primary objective of this study was to assess the effect of N95 FFR design, specifically the nose-clip shape and respirator style, on the fit factor (FF). The second aim was to assess how subject characteristics, specifically sex and race, affect the respirator fit. Additionally, the efficacy of N95 FFRs may degrade over prolonged use. The authors recently developed a portable device called the Exposure Protection Integrated Communicator (EPIC) that uses dual optical sensors to measure particle concentrations inside and outside the workplace and provide real-time monitoring of the workplace protection factor for FFRs. The EPIC's performance was evaluated by comparing its results with the reference PortaCount fit tester. Two cup-shaped FFRs with different noseclip shapes and six N95 FFRs with various styles were tested. Twenty human subjects were recruited to perform the Occupational Safety and Health Administration approved quantitative fit test (QNFT) protocol by running both PortaCount and EPIC devices in triplicate with each respirator. Results indicated that respirator FF was significantly affected by respirator design and subject facial characteristics. Caucasian males generally achieved the best fit, while Hispanic or Latino females had the poorest fit, highlighting the need for design improvements that account for differences in sex and race. Additionally, the comparison between EPIC and PortaCount demonstrated a strong proportional bias, with EPIC tending to report higher fit factor values, particularly under conditions where the respirator achieved higher fit factors. These findings underscore the need for cautious interpretation and potential calibration before EPIC can be reliably used in occupational settings.
More Related Videos
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Personal Protective Equipment