Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Beam-axis stability during gantry rotation measured with scintillation dosimetry
Shingo Ohira1,2, Takeshi Ohta3, Honoka Sugaya1
1Department of Radiological Sciences, Graduate School of Human Health Sciences, Tokyo Metropolitan University, Arakawa-ku, Tokyo, Japan.
This study shows that scintillation imaging accurately measures beam-axis stability during volumetric modulated arc therapy (VMAT) delivery. The results confirm submillimeter accuracy, supporting its use as a quality assurance (QA) tool.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Quality Assurance
Background:
- Volumetric modulated arc therapy (VMAT) demands high mechanical precision due to continuous gantry rotation.
- Ensuring beam-axis stability is critical for accurate dose delivery in VMAT.
Purpose of the Study:
- To investigate beam-axis stability during VMAT delivery.
- To evaluate scintillation-based imaging as a complementary quality assurance (QA) tool for VMAT.
Main Methods:
- Beam-axis deviations were measured using a scintillation plate dosimeter at 5 frames per second.
- Measurements were performed on a phantom with conformal arc beams under varying electric portal imaging device (EPID) and x-ray volume imaging (XVI) configurations.
- Short- and long-term reproducibility of beam-axis deviations were assessed, and three-dimensional displacements were calculated.
Main Results:
- Short-term reproducibility showed maximum standard deviations (SDs) below 0.11 mm for all tested energies.
- Long-term reproducibility exhibited maximum SDs up to 0.19 mm.
- Three-dimensional displacements between the radiation dose center and beam axis remained below 1 mm across all energies, with no significant impact from EPID/XVI configurations.
Conclusions:
- Scintillation-based imaging provides real-time, high-resolution assessment of beam-axis stability during VMAT.
- The observed deviations are consistently within submillimeter accuracy.
- This technique is suitable as a complementary QA method to conventional tests for VMAT.