Combined measurement uncertainty of HbA1c and glucose using ISO 20914:2019 and outcome-based simulation
Emine Feyza Yurt1, Medine Alpdemir1, Mehmet Şeneş1
1Departmant of Medical Biochemistry, Ankara Training and Research Hospital, Ankara, Türkiye.
Objectives:
This study aimed to calculate the measurement uncertainty (MU) of HbA1c and glucose according to ISO/TS 20914:2019 guideline and to evaluate using indirect outcome-based analytical performance specifications (APS).
Methods:
MU was estimated using internal quality control data collected between October 2023 and August 2025. A total of 13,270 and 13,824 IQC results for HbA1c and 1,329 and 1,208 results for glucose were analyzed for Level 1 and Level 2 controls, respectively. HbA1c was measured by capillary electrophoresis using two analyzers (12 capillaries each), and glucose was measured using the hexokinase method on three platforms. Maximum allowable uncertainty (MAU) targets were determined using the APS Simulator based on 2,000 patient samples with simultaneous HbA1c and fasting plasma glucose results. Clinical decision limits were defined as 5.7 % (39 mmol/mol) and 6.5 % (48 mmol/mol) for HbA1c, and 5.6 and 7 mmol/L for glucose. Overall agreement was used as the primary performance metric.
Results:
Relative expanded MU (Urel) ranged from 4.56 to 4.59 % for HbA1c expressed in %, 6.36-7.87 % for HbA1c expressed in mmol/mol, and 3.68-4.36 % for glucose. Glucose Urel values remained below the minimum MAU limit (5.1 %), whereas HbA1c values exceeded the corresponding limits (2.8 and 4.7 % for % and mmol/mol). Sensitivity and specificity ranged from 83.9 to 96 % and 93.4 to 99 % for glucose, and 85.4 to 93.4 % and 89 to 97.6 % for HbA1c.
Conclusions:
Glucose measurements met the minimum APS targets derived from the simulation model, whereas HbA1c measurements exceeded these limits, indicating higher MU and a potential impact on patient classification around clinical decision limits.
More Related Videos
12:02Automated Modular High Throughput Exopolysaccharide Screening Platform Coupled with Highly Sensitive Carbohydrate Fingerprint Analysis
Published on: April 11, 2016
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Related Concept Videos
Uncertainty: Overview
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Propagation of Uncertainty from Systematic Error
Uncertainty in Measurement: Reading Instruments
Hyperglycemia
Uncertainty in Measurement: Accuracy and Precision