Investigating the impact of key algorithm parameters and patient-specific factors on the accuracy of CT ventilation
Jeremy Lim1, John Kipritidis2, Jeremy T Booth2,3
1Image X Institute, Faculty of Medicine and Health, The University of Sydney, Sydney, Australia.
Background:
Computed Tomography Ventilation Imaging (CTVI) is an investigational technique that has its basis in functional lung avoidance radiotherapy. It offers a cost-effective and accessible alternative to nuclear medicine imaging by generating lung ventilation maps from 4DCT or paired inhale/exhale breath-hold CT (BHCT) scans. Despite over a decade of clinical validation, there is still no consensus on how algorithm parameters and patient-specific factors influence CTVI accuracy. Further research is needed to understand CTVI's sensitivity to these variables and to standardize its implementation for clinical use.
Purpose:
This study evaluates how key algorithm parameters and patient-specific factors affect the accuracy of CTVI.
Materials And Methods:
CT ventilation images were generated from BHCT scans and compared to Galligas PET ventilation scans. The VESPIR toolkit was used to compute ventilation based on deformable image registration (DIR) evaluation of volume change (CTVIJac) or change in Hounsfield Unit (HU) value (CTVIHU). CTVI accuracy was characterized as the voxel-wise Spearman correlation (rS) with Galligas PET. Algorithm parameters common to many CTVI implementations were investigated with a baseline determined from existing literature: lung segmentation threshold (-600 HU to -150 HU), DIR regularization parameter (λ = 0.05 to 100), and smoothing filter diameter (0 voxels to 9 voxels). Robust parameter ranges were defined as those yielding rS within 10% of the maximum cohort average observed through parameter variation, and no negative Jacobian values for the registration. Patient-specific lung volume and density metrics were also analyzed to explain inter-patient variability in CTVI accuracy.
Results:
The correlation between CTVI and Galligas PET was demonstrated to be robust within identified parameter ranges: lung segmentation threshold -600 HU to -150 HU for CTVIJac and CTVIHU, DIR regularization parameter (λ) 1.25 to 5 for CTVIJac and CTVIHU, and smoothing filter diameter 0 to 9 voxels for CTVIJac and 7 to 9 voxels for CTVIHU. No significant correlation was found between the accuracy of CTVIJac and any patient-specific lung volume or density parameters. Significant correlations were found between the accuracy of CTVIHU and the percentage change in lung volume during inspiration (r = 0.72, p < 0.01) and the lung volume in the exhale phase (r = -0.63, p < 0.01). The correlation between CTVIJac and CTVIHU was found to be strongly correlated to CTVI accuracy.
Conclusions:
CTVI accuracy was relatively stable across the range of parameter values tested with no strong indication of the need for patient-specific parameter sets. Patient-specific differences appear to be a driving factor for inter-patient variability in CTVI accuracy as parameter selection alone was insufficient to explain the variability. The strong association of CTVIJac and CTVIHU agreement and CTVI accuracy suggests that CTVIJac and CTVIHU agreement is a useful predictor of CTVI accuracy and quality metric for parameter optimization.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
09:08Acquiring Hyperpolarized 129Xe Magnetic Resonance Images of Lung Ventilation
Published on: November 21, 2023
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Imaging Studies for Cardiovascular System V: CT
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: