Excitation-Scanning Hyperspectral Imaging Microscopy to Efficiently Discriminate Fluorescence Signals
Published on: August 22, 2019
Rapid fluorescence excitation-emission matrix measurements using hybrid configuration of Michelson interferometer and
Abstract:
A technique for rapid fluorescence excitation-emission matrix (EEM) measurements is proposed, which leverages the high optical throughput and tunable resolution of a scanning Michelson interferometer and the fast acquisition of a grating spectrometer. In this technique, a broadband excitation light beam is modulated by a galvanometer-scanning Michelson interferometer, and the resulting interferometrically modulated light excites the sample. By synchronizing fluorescence detection with galvanometer scanning, fluorescence spectra corresponding to different optical path differences are separated. This configuration enables the tradeoff between the excitation module's spectral resolution and measurement speed, while enabling high throughput excitation light focused into a small spot on the sample. It is demonstrated that with excitation and emission spectral resolutions of 10 nm and 0.44 nm, respectively, the system can acquire the EEM of an 8.34 μM Rhodamine 6 G solution in ethanol in only 977 ms.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Fluorescence and Phosphorescence: Instrumentation