Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Real-time AI integration for MR to detect artifacts and guide pulse sequence adaptations.
Aaron T Gudmundson1, Zahra Shams2, Abdelrahman Gad2
1The Malone Center for Engineering in Healthcare, Johns Hopkins University, Baltimore, MD, USA.
Biorxiv : the Preprint Server for Biology
|May 18, 2026
Summary
This study introduces PEREGRINE, an AI-integrated MR pulse sequence that detects and corrects out-of-voxel (OOV) artifacts in real-time. The AI system successfully reduced OOV contamination, improving data quality in magnetic resonance spectroscopy (MRS).
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Artificial Intelligence (AI)
- Medical Spectroscopy
Background:
- Out-of-voxel (OOV) artifacts can compromise the quality of magnetic resonance spectroscopy (MRS) data.
- Real-time artifact detection and correction are crucial for improving diagnostic accuracy.
Purpose of the Study:
- To develop and evaluate a novel AI-integrated MR pulse sequence for real-time detection and correction of OOV artifacts.
- To prospectively update the crusher gradient scheme within the repetition time (TR) to minimize OOV contamination.
Main Methods:
- The PEREGRINE (Per Excitation Real-time Execution & Guided Responses with Integrated Neural-network Evaluation) system utilized convolutional autoencoders for OOV artifact detection.
- Scans were performed on healthy volunteers using edited MRS, comparing AI-off and AI-on conditions.
- The AI system triggered gradient scheme updates based on quantified OOV scores, iterating through 48 permutations.
Main Results:
- PEREGRINE provided real-time OOV scores and updated gradients within each 2-second TR.
- While overall OOV scores showed no difference in the 'Full' condition, the 'Dwell' condition demonstrated significantly lower OOV scores with AI-on compared to AI-off.
- Fit Quality Number (FQN) significantly improved in the AI-on scan, indicating enhanced data quality.
Conclusions:
- PEREGRINE successfully enabled real-time evaluation and reduction of OOV artifacts using an AI-integrated MR sequence.
- The system identified gradient modifications that effectively minimized OOV contamination, leading to improved MRS data quality.