The Diastolic Dominance Index: A Bedside Doppler Tool for Aortic Insufficiency in Continuous-Flow Left Ventricular
Sophie R Persaud1, Max Reisman2, Ryo Torii3
1From the Medical Scientist Training Program, Medical College of Wisconsin, Milwaukee, Wisconsin.
None:
Aortic insufficiency (AI) is a progressive complication of continuous-flow left ventricular assist device support and remains inconsistently graded during HeartMate 3 (HM3) therapy. Using a closed-loop cardiovascular simulation, we analyzed 2,312 physiologically screened HM3 profiles at pump speeds of 5,000 and 6,000 rpm to evaluate Doppler-based approaches for AI detection, defined by regurgitant fraction of greater than or equal to 30%. Established Doppler markers, including systolic-to-diastolic ratio (S/D), diastolic acceleration, and diastolic flow fraction (DFF), were compared with a derived Diastolic Dominance Index (DDI), which integrates systolic attenuation and diastolic predominance into a single metric. Systolic-to-diastolic ratio demonstrated strong discrimination (area under the receiver-operating characteristic curve [AUC]: 0.93 at 5,000 RPM; 0.86 at 6,000 RPM). Diastolic Dominance Index showed comparable performance (AUC: 0.87 and 0.83) and exhibited more stable discrimination across pump speeds, while providing a zero-centered index in which positive values indicate AI severity. Pump power alone was nondiagnostic, but in cases near the DDI decision boundary, it improved classification, achieving an AUC of 0.93 at 6,000 RPM. Simple rule-based combinations of DDI and power achieved greater than 90% accuracy at both speeds. By unifying established Doppler features into a single physiologic index and pairing it with an energetic adjudicator, this framework enables speed-robust, reproducible, and clinically interpretable AI grading during HM3 support. https://links.lww.com/ASAIO/B917.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Aortic Regurgitation I: Introduction
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...