Achieving lower radiation dose of cerebral computed tomography angiography: The role of high-concentration contrast
Wei Zheng1, Suying Wu1, Baoteng Zhang1
1Department of Radiology, The First Hospital of Putian City, Putian 351100, Fujian, China.
Objectives:
To investigate whether high-concentration iodine contrast medium (HICM) combined with low-dose radiation scanning protocols can further reduce the radiation dose of cerebral computed tomography angiography (CTA) scans without compromising image quality.
Materials And Methods:
The prospective study included 150 eligible patients who underwent cranial CTA examinations. The patients were randomly divided into three groups (A, B, and C). All patients received a total iodine dose of 21.6 g and underwent scanning at tube voltage of 80 kV. Each group included 50 patients. Different protocols were applied as below. Group A: Iomeprol 400, iodine delivery rate (IDR) 2.0 g/s, tube current-time product (TCTP) 90 mAs; Group B: Iopromide 370, IDR 1.48 g/s, TCTP 164 mAs; Group C: Iopromide 370, IDR 2.0 g/s, TCTP 90 mAs. Radiation doses (Computer Tomography Dose Index Volume, Dose-Length Product, Effective Dose) were recorded, and CT attenuation values, image noise, Signal-to-Noise Ratio (SNR), and Contrast-to-Noise Ratio (CNR) were measured.
Results:
There were no statistically significant differences in the SNR and CNR of intracranial vessels among groups A, B, and C. Group A and C showed higher CT attenuation values and image noise, but lower radiation dose (reduced by 46.53% and 45.62% compared with Group B). The injection flow rate was lower in Group A.
Conclusion:
A higher IDR is effective in reducing radiation dose. Moreover, higher iodine concentration (iomeprol 400) enables a lower injection flow rate to achieve a target IDR. A 46% reduction in radiation dose can be achieved using iomeprol 400 combined with a low-dose protocol, without compromising image quality, at an optimized injection flow rate.
Related Concept Videos
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI