Related Experiment Video
Updated: May 21, 2026

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Multicenter Validation of an AI-Based Snoring Model vs. Wearables for Diagnosing Sleep Apnea
Haoyan Guan1, Junyu Huang2, Danling Liu1
1Department of Otolaryngology Head and Neck Surgery, The Second People's Hospital of Shenzhen (the First Affiliated Hospital of Shenzhen University), Shenzhen, Guangdong, People's Republic of China.
Background:
Sleep apnea-hypopnea syndrome (SAHS) has high prevalence and cardiovascular/cerebrovascular comorbidities. Polysomnography (PSG) is the diagnostic gold standard but limited by complexity, cost, and accessibility; smart wearables are convenient but less accurate. This study validated an AI-based snoring-sound model and compared its diagnostic performance with a wearable, using PSG as reference.
Methods:
Adults with suspected SAHS (n = 134) were included and underwent overnight PSG, with simultaneous snoring sound and wearable signal recording. The snoring algorithm used short-time Fourier transform (STFT) spectrograms and a multiscale encoder-attention-decoder neural network for classification. Model performance was assessed using correlation analysis, intraclass correlation coefficient (ICC), Bland-Altman plots, and receiver operating characteristic (ROC) analyses. Accuracy, sensitivity, specificity, and area under the curve (AUC) values were computed for both models.
Results:
Disease severity was associated with variation in BMI (p < 0.001) and decreasing oxygen saturation (p < 0.001). The snoring model showed stronger correlation and higher agreement with PSG-derived AHI (r = 0.79, ICC = 0.753) than the wearable (r = 0.68). Although the snoring model yielded consistently higher AUCs, particularly in mild SAHS (AUC = 0.83 vs. 0.72, p = 0.07), the differences between methods were not statistically significant (all p > 0.05). Both models achieved excellent discriminative performance in severe SAHS (AUC ≥ 0.90).
Conclusion:
The AI-driven snoring-sound model demonstrated comparable overall performance and numerically better performance in certain settings, particularly in identifying mild SAHS.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
04:54IntelliSleepScorer, a Software Package with a Graphic User Interface for Mice Automated Sleep Stage Scoring
Published on: November 8, 2024