How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Comparison of Masimo O3 and INVOS 7100 cerebral oxygenation during immediate neonatal transition
Prathomwalee Sintupech1, Amonrat Anantho2, Kamonrat Songnok2
1Division of Neonatology, Department of Pediatrics, Faculty of Medicine Siriraj Hospital, Mahidol University, 2 Wanglang Road, Bangkok Noi, Bangkok, 10700, Thailand.
Abstract:
This study aims to compare CrSO2 measured simultaneously by Masimo O3 and INVOS during the first 15 min after birth in healthy term infants and to evaluate agreement and derived cerebral fractional tissue oxygen extraction (cFTOE). In this prospective observational study, 65 healthy term infants were monitored immediately after birth using Masimo O3 and INVOS 7100 applied simultaneously with randomized hemispheric placement. Pre-ductal oxygen saturation (SpO2) was recorded continuously. Minute-specific values were derived from aggregated measurements during the final 30 s of each minute. Paired comparisons were performed using Wilcoxon signed-rank test. Agreement was assessed using Bland-Altman analysis with regression evaluation for proportional bias. Both devices demonstrated a similar temporal increase in CrSO2 during the first 5 min, with relative stabilization thereafter. From minute 4 onward, Masimo O3 values were lower than INVOS, with significant differences at minutes 5 to 11 and min 14 (p < 0.05). Overall mean bias was - 2.8%, with 95% limits of agreement from + 18.0 to - 23.6%. Significant proportional bias was present (R2 = 0.11, p < 0.001). When CrSO2 was < 60%, mean bias was + 28.7% with wide limits of agreement. Masimo O3 derived cFTOE values were consistently higher at several time points. No significant hemispheric differences were detected.
Conclusion:
Masimo O3 and INVOS show similar temporal patterns but an oxygenation-dependent proportional bias. CrSO2 values are device-specific and not interchangeable, particularly at lower saturation levels during early transition.
What Is Known:
• Cerebral oxygenation during neonatal transition increases rapidly after birth and can be monitored using near-infrared spectroscopy. • Absolute values differ between devices, limiting comparability across NIRS systems.
What Is New:
• This study provides a simultaneous head-to-head comparison of Masimo O3 and INVOS during immediate neonataltransition, demonstrating oxygenation-dependent proportional bias. • CrSO2 values are device-specific and and should not be directly translated between monitoring platforms.
More Related Videos
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...