Related Experiment Video
Updated: May 22, 2026

A Three-Dimensional Digital Model for Early Diagnosis of Hepatic Fibrosis Based on Magnetic Resonance Elastography
Published on: July 21, 2023
Dynamic contrast-enhanced liver MRI in children: image quality comparison between GRASP-VIBE and CAIPIRINHA-VIBE
Simon Veldhoen1, Anne Pohrt2, Matthias Stephan Anders3
1Department of Radiology, Pediatric Radiology, Charité-Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany. simon.veldhoen@charite.de.
Objectives:
To compare image quality between GRASP-VIBE (free-breathing) and CAIPIRINHA-VIBE (breath-hold) sequences in dynamic liver MRI in a pediatric population.
Materials And Methods:
This retrospective study included 70 children who underwent dynamic liver MRI at 3 T with either GRASP-VIBE (n = 40) or CAIPIRINHA-VIBE (n = 40) acquisition. Ten patients underwent both imaging protocols at different time points. Image pairs were matched at four representative anatomical levels using vascular landmarks. Four dynamic contrast phases (pre-contrast, arterial, portal venous, delayed venous) at each anatomic level were independently reviewed by two radiologists, leading to a total of 640 comparisons per reader. Reader preference was assessed using a two-alternative forced-choice paradigm. Inter- and intra-reader agreement were analyzed using binomial tests and Cohen's κ.
Results:
GRASP-VIBE was preferred over CAIPIRINHA-VIBE across all contrast-enhanced phases of pediatric liver MRI: 82.2% [526/640] in the arterial, 83.1% [532/640] in the portal venous, 71.3% [456/640] in the delayed venous (all p < 0.001). No significant difference was observed for the pre-contrast phase with 59.4% [380/640] GRASP-VIBE preference (p = 0.45). Strongest GRASP-VIBE preferences were observed in older children and adolescents and in examinations performed without anesthesia. Across all anatomical levels, GRASP consistently outperformed Caipirinha-VIBE in contrast-enhanced phases. Inter-reader agreement was moderate in the arterial (κ = 0.51) and delayed venous (κ = 0.48) phases, and slight in pre-contrast (κ = 0.19) and portal venous (κ = 0.20) phases (intra-reader agreement, κ = 0.56, κ = 0.60, κ = 0.28 and κ = 0.42).
Conclusion:
GRASP-VIBE offered superior subjective image quality compared to CAIPIRINHA-VIBE in pediatric liver MRI, especially in early contrast phases. Its free-breathing approach might be particularly advantageous regarding image quality in children unable to perform breath-holds.
Key Points:
Question Does free-breathing GRASP-VIBE provide superior image quality compared to breath-hold CAIPIRINHA-VIBE for dynamic liver MRI in pediatric patients? Findings GRASP-VIBE was preferred across all contrast phases, delivering consistently higher-quality images than CAIPIRINHA-VIBE, particularly in children unable to perform breath-holds. Clinical relevance Using GRASP-VIBE enables high-quality dynamic pediatric liver MRI. Further research is needed to determine whether this sequence can reduce the need for sedation or general anesthesia in children.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies VII: Vascular Imaging