Ambient Sound Pressure Levels in the Bronchoscopy Suite: Contrasting Flexible, Rigid, and Robotic-Assisted
Rastko Rakočević1, Chelsea Dwyer1, Kam S Ho2
1Department of Medicine, Section of Interventional Pulmonology, Pulmonary Service.
Background:
Operating room noise can result in provider- and patient-related adverse outcomes. The recommended sound pressure threshold in medical settings is 35-45 dB(A). This study aimed to quantify sound pressure across different types of bronchoscopy procedures.
Methods:
Sound meters were placed in the bronchoscopy suite. Ambient sound pressure levels were measured continuously during flexible, rigid, and robotic bronchoscopy procedures. Procedures were classified as "quiet" or "loud" using the median equivalent continuous sound pressure level (Leq). A logistic regression model was used to identify associations between the type of bronchoscopy and "loud" or "quiet" procedures.
Results:
A total of 186 procedures were analyzed, of which 64% were flexible, 11% were rigid, and 25% were robotic. Across all procedures, the median Leq was 56.8 (55.1 to 58.1) dB(A). Leq was 58.2 (57.5 to 58.9), 59.4 (58.6 to 60.5), and 57.9 (57.5 to 58.4) dB(A) for flexible, rigid, and robotic bronchoscopy, respectively (P<0.0001). In comparison to flexible bronchoscopy, rigid bronchoscopy was more likely to be loud, while robotic bronchoscopy was less likely to be loud.
Conclusion:
This is the first study to quantify sound pressure levels during bronchoscopy procedures. Across all types of bronchoscopy procedures, sound pressure levels exceeded those recommended by guidelines for the medical setting and were considered a moderate degree of noise. Rigid bronchoscopy was associated with higher sound pressure levels, whereas robotic-assisted bronchoscopy was associated with lower sound pressure levels when compared with flexible bronchoscopy. Future studies are needed to explore the potential deleterious effects of bronchoscopy noise on operators and patients to inform preventive measures.
Related Concept Videos
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities:
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Physical Assessment of the Respiratory Tract III: Percussion
Percussion in Respiratory Assessment
Percussion evaluates underlying tissue composition with audible and tactile vibrations,...