Multi-Feature Adaptive Variational Mode Decomposition for Wearable ECG Devices
Zixin Chen1, Di Wu1, Yuanlin Nie1
1School of Electronic Information, Central South University, Changsha 410083, China.
Abstract:
To address the issue of motion artifact interference faced by wearable ECG monitoring devices in dynamic environments, this paper proposes an adaptive motion artifact removal framework based on improved Variational Mode Decomposition (VMD). By designing a parameter self-adjustment mechanism and a multi-feature fusion mode selection strategy, the algorithm's adaptability to non-stationary ECG signals and noise separation accuracy are enhanced. Experiments on the MIT-BIH Arrhythmia Database demonstrate that the improved VMD algorithm outperforms traditional wavelet transform, Recursive Least Squares (RLS), and conventional VMD methods in multiple performance metrics. Specifically, the signal-to-noise ratio (SNR) is improved by 5.17 dB, the Percentage Root Mean Squared Difference (PRD) is reduced to 49.13%, the correlation coefficient is increased to 0.88, and high real-time processing capability (Real-Time Processing Ratio, RTR = 22.5) is maintained, meeting the low-latency requirements of wearable devices. Moreover, case studies on pathological recordings (e.g., Wolff-Parkinson-White syndrome and third-degree atrioventricular block) reveal that the improved VMD better preserves clinically significant features such as delta waves and dissociated P waves. Furthermore, a downstream arrhythmia classification task using a CWT-CNN classifier achieves 91.67% accuracy on denoised heartbeats, which is 2.67 percentage points higher than that on raw noisy signals (89.00%), confirming the practical benefit of the proposed preprocessing for AI-based diagnosis. This study provides an effective processing solution for improving the signal quality of wearable ECG monitoring.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...