Evaluation of a Rapid Lateral Flow Assay for Coccidioidomycosis to Monitor Antibody Levels in Patients Using
Francisca J Grill1,2, Karen Pedersen3, Kenta Reilly4
1School of Life Sciences, Arizona State University, Tempe, AZ 85287, USA.
Abstract:
The diagnosis of coccidioidomycosis is often achieved serologically by the detection of antibodies against fungal antigens. While several serologic tests are available for coccidioidomycosis, all of them are performed in a laboratory setting causing delays in diagnosis and therapeutic intervention. Point-of-care testing offers the ability to provide a shorter time to result by avoiding specimen send-out, minimizing processing steps, and employing expeditious immunochemical techniques. A preliminary trial of a rapid anti-coccidioidal antibody lateral flow assay (LFA) using fingerstick blood was performed on 22 patients with coccidioidomycosis at the point of care during outpatient clinic visits. Patients were tested longitudinally over the course of one year. An LFA reader was implemented to provide an objective result by quantifying the intensity of the test line band. There was close qualitative concordance observed between positive LFA results with send-out immunodiffusion (89.5%) and complement fixation (78.4%) standard of care clinical laboratory assays. Additionally, the relationship between LFA test line density values and traditional complement fixation antibody titers was assessed.