Related Experiment Video
Updated: May 29, 2026

Determining Viral Disinfection Efficacy of Hot Water Laundering
Published on: June 21, 2022
Sterilization efficacy of a warm-air circulation system in a vaporized hydrogen peroxide sterilizer
Mi Suk Noh1, Su Yeon Lee2, Eun Gyu Lee2
1Bio&Medical Research Center, Bio Business Division, Korea Testing Certification Institute, Gunpo, Gyeonggi-do, Republic of Korea.
Background:
Vaporized hydrogen peroxide (VH2O2) sterilization is a widely recognized method for low-temperature sterilization of medical devices. To overcome limitations associated with residual moisture and maximize sterilization efficacy, the integration of a warm-air circulation system (WACS) is proposed as a robust solution.
Objective:
This study aimed to evaluate the effects of the WACS by comparing temperature uniformity and moisture removal with and without its operation. Additionally, the study assessed whether the WACS contributes to effective sterilization under conditions involving complex medical devices.
Methods:
Temperature variations within the sterilization chamber and medical devices were monitored using sensors. The impact of the WACS on sterilization efficacy was assessed by introducing artificial moisture to the loading packages in cycles performed with and without the system. Sterilization efficacy was assessed through tests with various items, including process challenge devices (PCDs), flexible and rigid lumens, and dual flexible endoscope dummies.
Results:
The WACS maintained a stable temperature of approximately 55 ~ 60 °C and effectively removed moisture throughout sterilization cycles, ensuring successful sterilization of all items.
Conclusion:
This study demonstrates that the WACS significantly improves sterilization effectiveness by ensuring uniform temperature and effective moisture removal. This additional system in the VH2O2 sterilizers will enhance the safety and sterilization efficacy of complex medical devices in healthcare settings.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Temperature
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Vaporization