Three-Dimensional Phase Resolved Functional Lung Magnetic Resonance Imaging
Published on: June 21, 2024
Synthetic phase image modulated by two-dimensional sinusoidal profile based on magnitude image using for magnetic
Zhaotong Li1, Junting Xi2, Jiahao Yang3
1Laboratory of Digital Medicine, Department of Medical Informatics, Medical School of Nantong University, Nantong 226001, China.
Background:
Magnetic resonance imaging (MRI) is a complex-valued technique incorporating magnitude and phase information, with phase images critical for susceptibility-weighted imaging and quantitative susceptibility mapping yet often absent in reconstruction. This study aimed to synthesize phase images from magnitude-only data via innovative phase modulation to reconstruct complete MR images.
Methods:
Synthetic phase information was generated using irregular phase modulation based on two-dimensional sinusoidal functions. Comprehensive assessments were performed to validate the synthetic phase performance. First, SVG (Synthetic Variable Gradient) and PSE (Phase-Shift Estimation) experiments were performed to evaluate the spatial heterogeneity of the synthesized phase structures. Second, five metrics, including signal-to-noise ratio (SNR), contrast, correlation, homogeneity, energy, evaluated synthetic phase image similarity to true phase image, the Wilcoxon rank-sum test and Bland-Altman analyses further validated the consistency and approximation between synthetic and true phase images. Final, compressed sensing (CS) and Dense-U-Dense Net (DUD-Net) were utilized to verify the feasibility of synthetic phase images for MRI reconstruction, with peak signal-to-noise ratio (PSNR), root mean square error (RMSE) and structural similarity index (SSIM) used for comparative evaluation.
Results:
The SVG and PSE results confirmed that the irregular phase provides more effective spatial heterogeneity than regular modulation. Statistical analyses demonstrated that the irregular phase achieved higher consistency with true phase images across five metrics including SNR, contrast, correlation, homogeneity, energy, with the Wilcoxon rank sum test (P < 0.05) and Bland Altman analyses further confirming its significant approximation to real physical phase. Regarding MRI reconstruction, the proposed irregular method demonstrated superior fidelity compared to regular methods. Quantitative validation showed that the irregular method achieved a DUD-Net PSNR of 34.90 ± 5.50 for magnitude reconstruction, which significantly outperformed the regular method and demonstrated comparable performance to the magnitude only reconstruction baseline. While irregular phase metrics were lower in high noise background areas, these have a competitive performance with the true phase when background noise was removed.
Conclusions:
Both CS and DUD-Net effectively reconstructed phase information, and phase modulation can provide reference phase data for establishing MR image reconstruction models without direct phase information. Notably, this heuristic synthetic phase serves as an approximation for reconstruction studies and cannot substitute true phase data for advanced physics-driven applications such as quantitative susceptibility mapping or susceptibility-weighted imaging.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse.
Graphical and Analytic Representation of Sinusoids
The first step is measuring the peak-to-peak value, which is twice the amplitude of the sinusoid. This provides information about the maximum voltage swing of the waveform.
Secondly, the period and angular frequency are determined. The period is the time taken for one complete cycle of the waveform, while...
Magnetic Resonance Imaging
Reconstruction of Signal using Interpolation