Quantitative 161Tb SPECT/CT imaging for dosimetry using a ring-based digital CZT camera
Eline Zoetelief1, Marcel Segbers2, Johannes Hofland3
1Department of Radiology & Nuclear Medicine, Erasmus MC Cancer Institute, Rotterdam, 3015 CN, The Netherlands. e.zoetelief@erasmusmc.nl.
Background:
161Tb-labeled compounds are emerging as promising alternatives to 177Lu-labeled compounds for radioligand therapy (RLT). The assessment of RLT safety and efficacy rely on accurate dosimetry, which necessitates quantitative activity measurement from multiple SPECT acquisitions. However, no studies to date have assessed 161Tb quantification using the ring-based cadmium-zinc-telluride (CZT) digital SPECT/CT systems, nor have optimal reconstruction parameters been established. This study aimed to optimize 161Tb SPECT reconstruction parameters for a CZT-based system.
Methods:
A phantom study was performed using the NEMA imaging quality phantom with the standard sphere set (diameters 10-37 mm), with the background prepared either with a 10.5:1 activity ratio or without activity. Imaging was conducted on a StarGuide (GE Healthcare) ring-based CZT digital SPECT/CT system using the manufacturer's recommended protocol. Scans were repeated over activity levels up to 2.2 GBq. Three reconstruction methods were evaluated: OSEM, Q.Clear without regularization term (QClearN) and Q.Clear with regularization term (QClearR), each tested with multiple iteration-subset combinations. To compare the reconstruction methods, recovery coefficient (RC), noise levels and spatial resolution were assessed.
Results:
RC values increased with sphere size and number of updates. At 864 updates, RC ranged from 0.56 to 0.95 for OSEM, 0.52-0.92 for QClearN, and 0.52-0.92 for QClearR. Noise levels, quantified by the coefficient of variation, in OSEM increased from 0.07 to 1.16 with increasing number of updates, while the noise levels of QClearN and QClearR ranged from 0.07 to 0.40 and 0.07 to 0.38, respectively. Spatial resolution was assessed using the full with at half maximum (FWHM) derived from matched-filter analysis of a digital phantom, with values of 7.9-17.0 mm for OSEM, 8.7-17.7 mm for QClearN, and 8.8-17.8 mm for QClearR. Gibbs artifact appeared in all methods, intensifying with higher number of updates.
Conclusions:
This study demonstrated the feasibility of quantitative 161Tb imaging using a ring-based CZT SPECT/CT system. Recommended number of updates are 48-96 for OSEM and approximately 200 for QClearN and QClearR. With these settings, 161Tb achieved higher RC values and improved spatial resolution compared to published 177Lu data, supporting its potential as a substitute for 177Lu for image-based dosimetry in RLT.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography