Related Experiment Video
Updated: Jun 2, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Safety and Feasibility of Pulsed Field Ablation With a Pentaspline Catheter in Patients With Extravascular ICD
Koumei Onuki1, Jun Hirokami1, Hidetaka Chinen1
1Department of Cardiology, Kokura Memorial Hospital, Kitakyushu, Japan.
Abstract:
Pulsed-field ablation (PFA) in patients with an implanted extravascular implantable cardioverter-defibrillator (EV-ICD) poses theoretical risks of electromagnetic interference or lead displacement. We report two cases of atrial fibrillation (AF) ablation using the FARAPULSE pentaspline catheter after EV-ICD implantation. In both patients, PFA was successfully performed without significant changes of the parameters/functions of the EV-ICDs. No lead dislodgement was observed on a fluoroscopy examination. These cases suggest that AF ablation using PFA can be safely performed in patients with EV-ICDs.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017