Phantom-based comparison of image quality in two digital PET/CT systems with different axial fields of view
Khomotso Marlene Legodi1,2, Milani Qebetu1,2,3, Kaluzi Banda1,2
1Nuclear Medicine Research Infrastructure (NuMeRI), Pretoria, South Africa.
Abstract:
Objective.Silicon photomultiplier (SiPM)-based positron emission tomography (PET)/ computed tomography (CT) scanners offer enhanced sensitivity and timing resolution compared with conventional detector designs. This study compared two SiPM PET/CT systems sharing the same technology platform but differing in axial field of view (AFOV) and system sensitivity, to evaluate their impact on image quality, lesion detectability, and partial-volume effects.Approach.A flangeless Esser PET phantom, filled with gallium-68 [68Ga] and containing hot cylinders (8-25 mm) and cold spheres (9.5-31.8 mm), was imaged on both systems. Acquisitions employed matched reconstruction protocols. The shorter AFOV system used a 5 minute scan, while the longer AFOV system used a 2.5 minute scan to assess performance under reduced acquisition time. Image quality was quantified using contrast recovery coefficient (CRC), contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR).Main results.The longer AFOV system exhibited higher CRC and SNR for medium-to-large hot cylinders despite the shorter acquisition (16 mm cylinder CRC: 53.18 vs 88.12; SNR: 4.11 vs 7.14; 25 mm cylinder CRC: 84.30 vs 92.42; SNR: 17.99 vs 30.08 for shorter vs longer AFOV systems). The shorter AFOV system performed better for the smallest hot cylinder (8 mm; CRC: 73.79 vs 64.09; CNR: 343.98 vs 258.99) and consistently achieved higher CRC and CNR for cold spheres, particularly at smaller diameters.Significance.The longer AFOV system provides superior contrast recovery and noise performance for medium-to-large lesions, even with reduced scan duration. Conversely, the shorter AFOV system offers advantages for very small hot and cold targets, illustrating the trade-off between sensitivity and spatial resolution. These system-specific characteristics should guide protocol optimization for different clinical imaging tasks.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT