Related Experiment Video
Updated: Jun 4, 2026

A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Comparative strain‑gauge analysis of Gamma3 and Gamma3 reconstruction nails in unstable basicervical femur fractures:
Jose Antonio Valle-Cruz1, Yaiza Lopiz1,2, Daniel Garríguez-Pérez3
1Department of Traumatology and Orthopaedic Surgery, Hospital Clínico San Carlos, Madrid, Spain.
Purpose:
Evidence supporting the mechanical superiority of the Gamma3 RC nail over the standard Gamma3 configuration remains limited. This pilot biomechanical study compared strain distribution and load-transfer behavior between both implants in a standardized unstable basicervical proximal femoral fracture model.
Methods:
Six fourth-generation composite femora were allocated into three groups: one intact reference specimen, three fractures stabilized with a standard Gamma3 nail, and three fractures stabilized with a Gamma3 RC nail incorporating a U-blade lag screw. Mechanical testing consisted of controlled low-load axial-oblique loading under standardized conditions. Four synchronized strain-gauge rosettes recorded cortical deformation at the proximal femur.
Results:
The Gamma3 RC construct demonstrated consistently lower strain magnitudes at the medial cortices and preserved a more linear strain-load response throughout the tested loading range compared with the standard Gamma3 configuration (Δε = 18.6 με and - 26.5 με). Differences between constructs became progressively more evident with increasing load (Δε = 180.9 με and - 62.9 με), particularly at the proximal medial cortex, where the standard Gamma3 showed greater strain concentration and early nonlinearity.
Conclusion:
In this exploratory synthetic-femur model, the Gamma3 RC nail demonstrated a mechanical trend toward more homogeneous load distribution and improved rotational control compared with the standard Gamma3 nail.