Endovascular repair of a type IA endoleak following ALTO endograft placement using a custom-made fenestrated device
Marc-Antoine Bety1,2, Stéphane Elkouri1,2, Laura M Drudi1,2
1Faculty of Medicine, Université de Montréal, Montreal, Quebec, Canada.
Abstract:
The ALTO endograft (Endologix, Inc) was developed to expand endovascular aneurysm repair eligibility. However, type IA endoleak remains a devastating complication associated with this device. Few reports describe management of endoleaks after ALTO implantation, and published cases are limited to open surgical explantation. We report a case of type IA endoleak following endovascular aneurysm repair using an ALTO device, which was successfully treated with a custom-made three-fenestration endograft with an integrated bifurcation. The procedure was performed with fusion imaging and carbon dioxide angiography. Technical success was achieved. One-year follow-up imaging demonstrated no residual proximal endoleak or aneurysm sac enlargement.