Related Experiment Video
Updated: Jun 9, 2026

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Core Temperature Estimation Using Wearable Earbud-Type Thermometer and Machine Learning During Light-Intensity
Naoto Fujii1,2, Akira Katagiri1, Kouichi Nakagome3
1Institute of Health and Sport Sciences, University of Tsukuba, Tsukuba, Japan.
None:
Continuous monitoring of core temperature is crucial for optimizing exercise performance, promoting health, and ensuring safety. Here, we evaluated the validity of a machine learning-based earbud-type core temperature sensor during exercise. At ambient temperatures of 10°C, 20°C, and 30°C, participants rested and then cycled for 25 min (90 W for males and 60 W for females), with an artificial wind of ~3.0 m/s directed at the face during the final 10 min of exercise. Throughout the experiment, two thermistors embedded in the earbud-type device monitored both ambient temperature and internal ear temperature. Additionally, an infrared temperature sensor integrated into the earbud-type thermistor measured internal ear temperature. Core temperature was measured through a machine learning-based approach, and esophageal temperature served as the reference. We created mixed-effects Bland-Altman plots for the relationship between estimated and esophageal temperature, and found that the bias was -0.003°C with a limit of agreement of -0.52°C-0.51°C. Root mean squared error and Pearson's r for the two-temperature relationship were 0.26°C and 0.81, and 76.9% of the data exhibited temperature differences within a margin of ≤ 0.3°C between the two measurements. Similar results were also observed with artificial wind. Including infrared temperature measurements did not enhance the aforementioned variables. We show that a machine learning-based wearable earbud-type thermometer can validly estimate core temperature in exercising individuals across 10°C-30°C indoor ambient temperatures with and without exposure to a wind speed of ~3.0 m/s. However, further improvement of the algorithm is needed to enhance estimation accuracy.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...