How Successful are Meniscus Repair All-Inside Implants during Deployment?
Emma L Flanigan1, Sonu Bae1, Tyler Barker2
1Sports Medicine Research Institute, The Ohio State University Wexner Medical Center, Ohio, United States, Columbus.
The Journal of Knee Surgery
|June 9, 2026
Summary
Arthroscopic all-inside meniscal repair demonstrates a low overall implant failure rate of 1.03%. Specific implants showed higher waste, warranting further investigation by surgeons and industry.
Area of Science:
- Orthopaedic Surgery
- Biomedical Engineering
- Sports Medicine
Background:
- Meniscal tears are common knee injuries requiring surgical intervention.
- Arthroscopic all-inside meniscal repair is a prevalent surgical technique.
- Implant performance is critical for successful meniscal repair outcomes.
Purpose of the Study:
- To determine the percentage of implants successfully deployed during arthroscopic all-inside meniscal repair.
- To assess the intra-operative failure or waste rate of various meniscal repair implants.
- To identify specific implants with higher failure rates in clinical practice.
Main Methods:
- Retrospective data query of meniscus repair procedures (CPT codes 29882, 29883) from June 2020 to June 2023.
- Analysis of implant usage and intra-operative failure data from a single institution.
- Calculation of implant success rate by dividing wasted implants by total implants used.
Main Results:
- A total of 1,026 patients and 3,867 meniscal implants were analyzed.
- The overall implant failure or waste rate was low at 1.03% (40 out of 3,867 implants).
- Specific implants exhibited higher failure rates: JuggerStitchTM Curved (9.38%), NOVOSTITCHTM Cartridge 0 Suture (2.44%), and TRUESPANTM 12 Degrees (2.11%).
Conclusions:
- The overall failure rate for implants used in arthroscopic all-inside meniscal repair is low.
- Certain implants demonstrate higher waste rates, necessitating attention from manufacturers and surgeons.
- Further research into implant failure rates and associated costs in meniscal repair is recommended.