Usability and Accuracy of a Novel Automatic Bladder Diary Device (Uro-diary Checker)
Yoshinaga Okumura1, Yoshie Amaya2, Sou Inamura1
1Department of Urology, University of Fukui, Fukui, Japan.
Objective:
To evaluate the usability and accuracy of a newly developed automated bladder diary (BD) device, the Uro-diary Checker (UDC), in comparison with a conventional handwritten BD.
Methods:
In this prospective, multi-institutional observational study, 40 patients with lower urinary tract symptoms (LUTS) were recruited between April 2023 and October 2024. Participants recorded their voiding patterns using a handwritten BD for 3 days and the UDC for the next 3 days. The UDC automatically recorded voiding time and volume when a urine cup was placed on the device, and the data were transferred via Bluetooth to a computer. The primary endpoint was usability, assessed using the System Usability Scale (SUS), a validated 10-item Likert-scale questionnaire.
Results:
Of the 40 enrolled patients, 31 successfully completed records using both methods and were included in the final analysis. The mean patient age was 72.6 ± 9.8 years; 24 were male and 7 were female. The SUS score was significantly higher for the UDC (72.9 ± 19.3) than for the BD (60.6 ± 19.5; P < .01), exceeding the benchmark score of 68 and indicating good usability. UDC-recorded urine volumes were highly consistent with BD measurements for both 24-hour urine volume (r = 0.86, P < .01) and average voided volume (r = 0.77, P < .01).
Conclusion:
The UDC demonstrated significantly greater usability than the conventional BD while maintaining comparable accuracy for urine volume measurements, suggesting its potential to improve adherence and efficiency in LUTS evaluation.
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Imaging Studies VI: Voiding Cystourethrography and Cystography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Automated Microbial Diagnostics