Related Experiment Video
Updated: Jun 11, 2026

Control of Eating Behavior Using a Novel Feedback System
Published on: May 8, 2018
Comparing Automated Lyumjev® Delivery with Carbohydrate Counting, Qualitative Meal-Size Estimation, and Fully
Carolyne Schumacher1, Michael A Tsoukas2,3, Laurent Legault4
1Department of Experimental Medicine, McGill University, Montreal, Quebec, Canada.
Objective:
Determining prandial insulin needs by carbohydrate counting is a challenging task for many people with type 1 diabetes (PwD). We developed a novel automated insulin delivery (AID) system using ultrarapid lispro insulin (Lyumjev®), replacing carbohydrate counting with qualitative meal-size estimation and meal detection.
Research Design And Methods:
We performed a randomized crossover pilot trial comparing 3 weeks of AID with carbohydrate counting (AID-count) with (i) AID with qualitative meal-size estimation (i.e., defining a meal as low, medium, high, or very high in carbohydrate content; AID-estimate) and (ii) AID with meal detection (i.e., fully closed-loop; AID-fully) in adults with type 1 diabetes (T1D). The trial was registered (NCT06021158).
Results:
Twelve participants completed the study (eight females; mean [SD]: age 38 [16] years, HbA1c 7.0% [0.7], total daily insulin dose 0.64 [0.22] U/[kg·d]). There were no differences greater than 5% in the overall time in range between AID-estimate and AID-fully compared with AID-count, and thus the differences were not considered clinically significant based on the international consensus on time in range (AID-count vs. AID-estimate: 75.4% [16.3] vs. 72.9% [13.2], P = 0.191; AID-count vs. AID-fully: 75.4% [16.3] vs. 71.0% [10.6], P = 0.192). The differences in time spent in hyperglycemia were not statistically significant, and the times spent in hypoglycemia were low and not different between interventions. There were no episodes of severe hypoglycemia or diabetic ketoacidosis in any intervention.
Conclusions:
In this pilot study, the AID system using Lyumjev simplified or alleviated carbohydrate counting without degrading glucose control more than 5%. Larger studies are warranted.
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into rapid-acting...
Type II Diabetes I: Introduction
Type I Diabetes I: Introduction
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...