Optimization of X-ray acquisition parameters for reduced radiation exposure while maintaining 2D-DSA image quality
Charyl Jia Qi Yap1, Byung Jun Kim2, Muhammad Syadad Bin Sulaiman3
1Department of Vascular Surgery, Singapore General Hospital, Singapore, Singapore.
Background:
Achieving optimal image quality with minimal radiation exposure allows operators to accurately assess and treat target lesions during percutaneous transluminal angioplasty (PTA) without being subjected to increased radiation risk. Depending on clinical needs, imaging protocols are optimized and evaluated. This study aims to compare two imaging protocols-conventional dose protocol (CDP) and low dose protocol (LDP)-to evaluate the impact of fine-tuned X-ray acquisition parameters (detector entrance dose, copper filtration, tube voltage) and post-processing parameters (edge enhancement, window contrast).
Methods:
Thirty patients who underwent PTA between February to April 2021 were retrospectively reviewed. A representative two-dimensional digital subtraction angiography (2D-DSA) image was selected for both image quality and radiation dose comparison. Radiation dose was assessed using dose area product per frame (DAP/frame) and air kerma per frame (AK/frame) as normalized metrics to exclude effect of procedural duration. Each image was assessed quantitatively by measuring contrast-to-noise ratio (CNR) and qualitatively by two experienced vascular surgeons with a 5-point Likert scale.
Results:
Radiation dose with LDP showed a significant reduction in DAP/frame (2.49±1.65 µGy·m2 in CDP vs. 0.48±0.42 µGy·m2 in LDP) and AK/frame (0.10±0.07 µGy in CDP vs. 0.02±0.02 µGy in LDP). CNR was higher with LDP than with CDP (0.83±0.81 in CDP vs. 2.99±2.44 in LDP). Qualitative image quality assessment demonstrated no significant differences between protocols across all criteria (P>0.05).
Conclusions:
Parametric optimization of X-ray acquisition protocols in angiography system can potentially reduce radiation exposure without compromising image quality.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System V: CT