Related Experiment Video
Updated: Jun 13, 2026

Probe-based Confocal Laser Endomicroscopy of the Urinary Tract: The Technique
Published on: January 10, 2013
Study of Bladder Cancer Detection in Standard White Light Versus AI-Supported Endoscopy-02 (RAISE-02)-A Randomized
Peter Blak Hjort1,2, Katharina Skovhus1,2, Jørgen Bjerggaard Jensen1,3
1Department of Clinical Medicine, Aarhus University, Palle Juul-Jensens Blvd. 82, 8200 Aarhus N, Denmark.
Background:
White light cystoscopy (WLC) is the gold standard for bladder cancer (BC) detection and surveillance, but has limited sensitivity, particularly for small and flat lesions. We evaluated the diagnostic performance and safety of a real-time artificial intelligence (AI)-assisted support tool (CystoAID©) for bladder cancer detection.
Methods:
This single-center, randomized, controlled non-inferiority trial included 64 patients undergoing transurethral resection of bladder tumor or laser fulguration for suspected primary or recurrent BC. Patients were randomized 1:1 to WLC alone or WLC followed by AI-assisted cystoscopy. The primary outcome was per-lesion sensitivity of CystoAID vs. WLC within the intervention group. The reference standard was histopathology or clinical evaluation. The predefined non-inferiority margin was -5%. Secondary outcomes included sensitivity for lesions ≤ 5 mm, false positives, procedural duration, and safety (30-day adverse events).
Results:
A total of 142 lesions were identified (84 intervention, 58 control). Sensitivity was 96.2% (95% CI 87.0-99.5) for CystoAID and 88.7% (95% CI 77.0-95.7) for WLC (difference 7.5%; 95% CI -2.7 to 17.8), demonstrating non-inferiority. In a sub-analysis for lesions ≤ 5 mm (n = 25), sensitivity was 100% (95% CI 86.3-100) vs. 80% (95% CI 59.3-93.2). AI assistance added a median of 2.1 min without further affecting workflow. Adverse event rates were low and similar between groups.
Conclusions:
CystoAID demonstrated non-inferior sensitivity to WLC and could be safely integrated into the routine clinical workflow. Larger studies with improved representation of flat lesions are warranted, as well as to assess whether CystoAID can achieve superior sensitivity compared with standard WLC.
Related Concept Videos
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Urologic Endoscopic Procedure: Cystoscopic Examination
Imaging Studies VI: Voiding Cystourethrography and Cystography
Imaging Studies II: Ultrasonography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography