Purpose:
Accurate assessment of the pressure gradient ( ) across aortic coarctation (CoA) is critical for determining disease severity and the need for intervention. Current non-invasive methods are unreliable, while invasive catheterization remains the clinical gold standard. This study evaluates a novel MRI acquisition strategy, 4D-FlowP, that simultaneously encodes blood velocity and acceleration to enable reliable non-invasive pressure gradient mapping in CoA.
Methods:
Patient-specific compliant aortic phantoms were created from clinical MRI data of two patients with CoA. Additional geometries were synthetically generated by increasing stenosis severity. Phantoms were studied in an MRI-compatible flow loop under physiologically realistic flow and pressure conditions. Pressure gradients were estimated using conventional 4D-Flow MRI, 4D-FlowP, and fluid-structure interaction (FSI) simulations. Results were compared against ground-truth catheter-based measurements across multiple flow rates and stenosis severities.
Results:
Conventional 4D-Flow consistently underestimated (slope = 0.63, ) relative to catheter measurements. In contrast, 4D-FlowP demonstrated substantially improved agreement (slope = 0.95, ). FSI simulations showed the highest overall agreement with catheter-derived (slope = 1.14, ). Scan times for 4D-FlowP were comparable to 4D-Flow (26 vs. 24 minutes).
Conclusion: