Radiographic repeat analysis in four public hospital radiology departments in the Souss-Massa region, Morocco: A
A El Ouaridi1, A Ben Lachger1, J Elkhalladi2
1Hassan First University of Settat, High Institute of Health Sciences, Laboratory of Sciences and Engineering of Biomedicals, Biophysics and Health, Settat, Morocco.
Introduction:
This study presents a radiographic repeat analysis conducted in public hospitals in the Souss-Massa region of Morocco. It aimed to assess repeat rates, identify the most frequent causes of image repetition and the most affected anatomical regions, and examine variations in recurrence rates.
Methods:
A cross-sectional study was conducted during March and April 2025 in four public hospitals in the Souss-Massa region of Morocco. The number of acquired and repeated images was recorded using a standardized observation grid, supplemented by repeated image extraction for retrospective analysis. Descriptive statistics and chi-square tests were used to assess repeat rates, associated causes, anatomical regions involved, and variations by hospital and day of the week.
Results:
Of the 15,493 radiographic images analyzed, 1063 were repeated, resulting in an overall repeat rate of 6.86%, which falls within the acceptable range recommended by the American Association of Physicists in Medicine (AAPM). The highest specific repeat rate was observed at Hospital 2 (11.94%), followed by Hospital 1 (4.63%), Hospital 4 (4.33%), and Hospital 3 (3.92%). The most frequently repeated examinations were chest, pelvis, abdomen, lumbar spine, and shoulder. Across all examination types, the main causes included positioning errors, metallic artifacts, collimation errors, and under- or over-exposure. Daily variability in repeat rates was also observed.
Conclusion:
Although the repeat rate was within acceptable limits, indicating reasonably adequate radiographic image production, variations between hospitals and the predominance of preventable professional errors highlight the need for improved technical consistency.
Implications For Practice:
Implementing standardized protocols, extending repeat-rate assessment and regular monitoring, establishing a reference threshold, improving understanding of human and organizational factors, and harmonizing image acceptance criteria would help reduce repetition and strengthen consistency in assessments across professionals.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging