Limitations in diagnostics and quantification of small lesions with low uptake in the clinical context of prostate
Maike E Lindemann1, Walter Jentzen2,3, Alina Küper2,3
1High-Field and Hybrid MR Imaging, University Hospital Essen, University of Duisburg-Essen, Essen, Germany. maike.lindemann@uk-essen.de.
Purpose:
The aim of this study was to investigate the limits of diagnostics and therapy planning for patients with prostate cancer using non-time-of-flight 18F/68Ga-PSMA PET/MRI under clinically challenging imaging conditions with small lesion sizes and low uptake. Lesion detectability and quantification accuracy were evaluated for different acquisition and reconstruction parameters in a systematic phantom study and subsequent on patient data.
Methods:
PET/MRI measurements were performed using a small lesion NEMA phantom. PET data were acquired for nine different activity concentrations (AC). Data of a longer single-bed protocol in the pelvis or a shorter whole-body protocol were reconstructed using relative or absolute scatter correction (SC). PET images were analysed considering a ± 25% deviation range between imaged and true AC as acceptable. Thirteen PSMA-PET/MRI patients with primary lesions or lymph node metastasis < 12 mm in the pelvis were included in this study. The presence of the halo artefact was evaluated in six 18F-PSMA and seven 68Ga-PSMA PET/MRI patients. For 21 lesions (diameter 6.4-12.3 mm) in total, the AC was quantified.
Results:
For both radiotracers, the 9.7 mm sphere was still visible at 0.16 kBq/mL with emission times > 40 min. The 3.7 mm sphere was only detectable at 22 kBq/mL with emission times > 4 min. All spheres ≥ 6.5 mm provide acceptable quantification at an AC of 1.32 kBq/mL for 18F PET/MRI protocols of ≥ 12 min and 2.75 kBq/mL for 68Ga PET/MRI protocols. In phantom data, no halo artefact was observable and different SC methods had no impact on quantification. 4/6 18F-PSMA patients and 7/7 68Ga-PSMA patients showed a halo around the bladder using relative SC, which could be reduced in all patients using absolute SC. Comparing the minimum quantifiable AC (MQAC) from in the phantom study as a threshold to the patient data, all lesions provided acceptable quantification with values > MQAC (AC 3.3-108.5 kBq/mL) for equal reconstruction and acquisition parameters.
Conclusions:
The results demonstrated that the detection of lesions in the sub-centimetre range and a reliable quantification of 18F/68Ga-PSMA uptake using standard acquisition and reconstruction parameters within clinical PET/MRI protocols is possible. This allows for an individual assessment of potential therapy options for each patient.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
06:08A Cognitive Fusion-guided Prostate Biopsy Using Multiparametric Magnetic Resonance Imaging and Transrectal Ultrasound
Published on: March 21, 2025
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...