Erythrocyte Sedimentation Rate: A Physics-Driven Characterization in a Medical Context
Published on: March 24, 2023
Rapid Measurement of Erythrocyte Sedimentation Rate Using a Tube Positioned at a 45° Angle Compared With the
Francis Agyei Amponsah1, Benedicta Serwaa Obeng1, Otchere Addai-Mensah1
1Department of Medical Diagnostics, Faculty of Allied Health Sciences, Kwame Nkrumah University of Science and Technology, Kumasi, Ghana, knust.edu.gh.
Biomed Research International
|June 19, 2026
Summary
Reducing erythrocyte sedimentation rate (ESR) test time is crucial for clinical workflows. Positioning the Westergren tube at a 45° angle significantly shortens the ESR test duration to 8 minutes while maintaining accuracy.
Area of Science:
- Hematology
- Clinical Chemistry
Background:
- The erythrocyte sedimentation rate (ESR) is a vital hematological test for detecting inflammation.
- It is cost-effective and widely used in resource-limited settings like Ghana.
- The conventional 1-hour Westergren method can delay clinical results.
Purpose of the Study:
- To investigate a modified Westergren method for faster ESR results.
- To reduce the ESR test turnaround time without compromising accuracy.
- To evaluate the feasibility of a 45° tube positioning for ESR estimation.
Main Methods:
- An experimental analytical method comparison study involving 100 participants (50 pregnant, 50 healthy).
- Venous blood samples were collected and mixed with trisodium citrate anticoagulant (4:1 ratio).
- Two Westergren-Katz tubes per participant were used: one vertical (90°) and one at a 45° angle.
Main Results:
- ESR results from the 6th to 10th minute at 45° correlated with 1-hour results at 90° (p > 0.05).
- Bland-Altman analysis indicated optimal agreement at the 8th minute with 45° positioning.
- The 45° method showed minimal bias compared to the standard 90° method.
Conclusions:
- The ESR test turnaround time can be reduced to 8 minutes.
- Positioning the Westergren tube at a 45° angle is a viable method for rapid ESR determination.
- This modification offers a faster alternative for ESR testing in clinical settings.