A scaffold-degradable single-rivet ASD occluder with amplatzer-like design: a multicenter randomized controlled trial
Liqing Zhao1, Jiawei Gao1, Xin Zhou1
1Department of Pediatric Cardiology Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Cardiovascular Intervention and Therapeutics
|June 19, 2026
Summary
The novel biodegradable Pansy occluder demonstrated 100% closure success in atrial septal defect (ASD) patients at 12 months, showing promising safety and efficacy. Longer-term data will confirm durability.
Area of Science:
- Cardiology
- Biomaterials Science
- Medical Devices
Background:
- Metallic atrial septal defect (ASD) occluders pose mid- to long-term complication risks.
- A novel biodegradable Pansy occluder using a polydioxanone (PDO) framework was developed to mitigate these risks.
Purpose of the Study:
- To evaluate the effectiveness and safety of the biodegradable Pansy ASD occluder compared to traditional metallic occluders.
- To assess the closure success rate at six months and twelve months post-implantation.
Main Methods:
- A prospective, multicenter, randomized controlled trial involving 112 adult and pediatric patients.
- Patients were randomized into two groups: Pansy occluder (experimental) and metallic occluder (control).
- Primary efficacy endpoint was the closure success rate at six months post-operation.
Main Results:
- 100% closure success rate at six months and twelve months in both the Pansy occluder and control groups.
- No residual shunt greater than 5 mm was detected in any patient during follow-up.
- One patient in the experimental group required explantation due to abnormal tissue hyperplasia; pathological analysis showed PDO framework degradation and endothelial coverage.
Conclusions:
- The biodegradable Pansy ASD occluder demonstrates excellent safety and efficacy at twelve months post-implantation.
- The PDO framework shows promising degradation characteristics with smooth endothelialization.
- Longer-term follow-up is necessary to fully establish the durability of the Pansy occluder.