Validation of a remote monitoring blood pressure device in a general population cohort
Matthew M Moore1, Ayamo Oben1,2, Paul Muntner1,3
1Center for Women's Reproductive Health, University of Alabama at Birmingham, Birmingham, AL, United States.
Background:
Remote blood pressure (BP) monitoring has emerged as a promising solution to overcome barriers to hypertension care access, yet validation of devices is lacking.
Methods:
We conducted the validation of the Ideal Life BP Manager BP device, a remote BP monitoring device, using the AAMI/ESH/ISO protocol. From June 2021 to May 2022, participants were screened and systolic BP (SBP) and diastolic BP (DBP) measurements were conducted by trained staff using the Ideal Life BP Manager and a mercury sphygmomanometer. Statistical analyses were performed to assess agreement of SBP and DBP.
Results:
Of the 90 individuals screened, 79 were included, with a mean age of 50.9 years (standard deviation [SD] 15.6) and the majority being female (68.8%). The mean ± SD differences between the Ideal Life BP Manager and mercury sphygmomanometer measurements were 2.20 ± 7.37 mm Hg for SBP and 1.49 ± 7.31 mm Hg for DBP, satisfying AAMI/ESH/ISO protocol criterion 1. Criterion 2 was also met, with participant-level standard deviations of the mean paired differences (6.54 mm Hg for SBP, 6.30 mm Hg for DBP) falling below the protocol-specified thresholds. Bland-Altman plots demonstrated limits of agreement of -12.0 to 16.9 mm Hg for systolic blood pressure and -12.8 to 15.9 mm Hg for diastolic blood pressure.
Conclusions:
The Ideal Life BP Manager met the AAMI/ESH/ISO validation criteria.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure