Analytical Evaluation of a Point-of-Care Platform for Glycated Hemoglobin and N-Terminal Pro-B-Type Natriuretic

Perrin Ngougni Pokem1,2, Ali Khatib3,2, Iman Azariouh Kaddouri4,5

  • 1Department of Clinical Biochemistry, Cliniques Universitaires St-Luc and Universite Catholique de Louvain, Brussels, Belgium.

Cardiology Research
|June 22, 2026
PubMed

Insights

The AFIAS-3 point-of-care testing platform shows excellent accuracy for glycated hemoglobin (HbA1c) and acceptable accuracy for N-terminal pro-B-type natriuretic peptide (NT-proBNP). This dual biomarker POCT supports decentralized cardiometabolic diagnostics.

Area of Science:

  • Cardiovascular diagnostics
  • Diabetes management
  • Point-of-care testing

Background:

  • Heart failure is a common diabetes complication, necessitating integrated cardiometabolic assessment.
  • Glycated hemoglobin (HbA1c) and NT-proBNP are key biomarkers for glycemic control and myocardial stress, respectively.
  • Point-of-care testing (POCT) offers potential for decentralized diagnostics and reduced turnaround times.

Purpose of the Study:

  • To evaluate the analytical performance of the AFIAS-3 POCT fluorescence immunoassay platform.
  • To compare HbA1c and NT-proBNP measurements with established central laboratory methods.

Main Methods:

  • A single-center verification study using 50 patient samples per analyte.
  • Comparison of AFIAS-3 HbA1c with HPLC (Tosoh G8) and NT-proBNP with electrochemiluminescence immunoassay (cobas 8000).
  • Assessment of analytical precision, accuracy, method agreement (Passing-Bablok, Bland-Altman, Spearman), and clinical threshold agreement (Cohen's kappa).

Main Results:

  • AFIAS-3 demonstrated excellent precision for HbA1c (CVs 1.9%, 1.8%) with strong correlation (r=0.98) and minimal bias (-0.03%).
  • Excellent agreement for HbA1c at the 6.5% diagnostic threshold (κ=0.92).
  • High correlation for NT-proBNP (r=0.99) with substantial agreement at clinical thresholds, despite observed proportional bias at higher concentrations (CVs 5.3%, 15.4%; slope 1.24).

Conclusions:

  • The AFIAS-3 POCT platform shows excellent analytical agreement for HbA1c.
  • Acceptable agreement for NT-proBNP at clinically relevant thresholds supports its use in decentralized cardiometabolic diagnostics.
  • NT-proBNP results from AFIAS-3 at higher concentrations require cautious interpretation.
Abstract