Biosensor for Detection of Antibiotic Resistant Staphylococcus Bacteria
Published on: May 8, 2013
Is Teicoplanin Resistance Detected By The Automated System In Staphylococcus spp Strains Reliable?
Salim Yakut1, Fadile Yildiz Zeyrek2
1Faculty of Medicine, Department of Medical Microbiology, Harran University; 21salimyakut21@gmail.com.
Automated systems are unreliable for detecting teicoplanin resistance in Staphylococcus isolates. Broth microdilution (BMD) is recommended for confirming resistance, especially for methicillin-resistant Staphylococcus aureus (MRSA) and coagulase-negative Staphylococcus (CoNS).
Area of Science:
- Microbiology
- Infectious Diseases
- Antimicrobial Resistance
Background:
- High rates of methicillin resistance observed in Staphylococcus aureus (71.7%) and all coagulase-negative Staphylococcus (CoNS) isolates.
- Teicoplanin is a critical last-resort antibiotic for treating infections caused by resistant strains like MRSA and methicillin-resistant CoNS.
Purpose of the Study:
- To evaluate the accuracy of automated susceptibility testing systems for teicoplanin resistance detection.
- To compare automated testing with broth microdilution (BMD) and gradient test methods for teicoplanin susceptibility.
- To investigate teicoplanin resistance patterns in S. aureus and CoNS isolates.
Main Methods:
- Collected 467 S. aureus and 66 CoNS isolates between June 2022 and December 2024.
- Performed teicoplanin susceptibility testing using an automated system, BMD, and a gradient test.
- Identified methicillin resistance using standard methods.
Main Results:
- Automated systems incorrectly identified 13 S. aureus and 4 CoNS isolates as teicoplanin-resistant.
- All 13 S. aureus isolates were susceptible by gradient test and BMD.
- BMD confirmed teicoplanin resistance in 4 CoNS isolates (1 S. capitis, 3 S. haemolyticus) from blood cultures.
Conclusions:
- Automated systems are unreliable for detecting teicoplanin resistance in Staphylococcus species.
- Broth microdilution (BMD) is essential for confirming teicoplanin resistance, particularly in CoNS.
- Close monitoring of teicoplanin susceptibility is crucial due to its role in treating resistant bacterial infections.