A single site, multi-operator precision study for second-generation HR-pQCT
Joshua R Stapleton1, Nina Z Heilmann2, Kristin Cattell1
1Department of Biomedical Engineering, Wake Forest University School of Medicine, Winston-Salem, NC 27101, United States.
JBMR Plus
|June 23, 2026
Summary
High-resolution peripheral quantitative computed tomography (HR-pQCT) offers precise bone measurements. Standardized technologist training ensures consistent scan-rescan precision across operators, crucial for longitudinal bone studies.
Area of Science:
- Orthopedics and Bone Imaging
- Medical Physics
- Radiological Sciences
Background:
- High-resolution peripheral quantitative computed tomography (HR-pQCT) is increasingly used for bone dynamics research.
- Existing HR-pQCT precision data often rely on older technology and limited operator/population diversity.
- Ensuring reproducibility is vital for longitudinal HR-pQCT studies.
Purpose of the Study:
- To investigate the influence of operator characteristics and participant demographics on HR-pQCT measurement precision.
- To establish reference values for scan-rescan precision in a standardized academic imaging center.
- To assess the impact of technologist variability on HR-pQCT outcomes.
Main Methods:
- 45 adult participants underwent same-day repeat HR-pQCT scans of the distal tibia and radius.
- Three imaging technologists acquired 15 scan pairs each, with standardized post-processing.
- Statistical modeling analyzed the effects of technologists and participant demographics (age, sex, BMI).
Main Results:
- Root-mean-square coefficients of variation were ≤1.3% for bone mineral density and <2% for area measures.
- Trabecular microarchitecture precision ranged from 1.14% to 2.82%.
- Cortical thickness precision was ≤1.99%, while cortical porosity precision was higher (20.90-24.66%).
- No significant differences in precision were found based on technologist or participant demographics.
- Image quality scores did not impact precision outcomes.
Conclusions:
- Standardized technologist training yields consistent HR-pQCT scan-rescan precision across operators.
- These findings provide reference values for least significant change in HR-pQCT parameters.
- The study supports the reliability of HR-pQCT for longitudinal bone research with standardized protocols.