Selecting Multiple Biomarker Subsets with Similarly Effective Binary Classification Performances
Published on: October 11, 2018
An improved approach to modelling patient reclassification: HbA1c as an example.
Christopher-John Lancaster Farrell1, Tze Ping Loh2, Tony Badrick3
1NSW Health Pathology, Department of Chemical Pathology, Royal North Shore Hospital, Sydney, Australia.
External quality assurance (EQA) programs assess in vitro diagnostic (IVD) devices. A new incremental reclassification rate (IRR) method found most HbA1c IVD devices perform well for diabetes diagnosis and monitoring.
Area of Science:
- Clinical Chemistry
- In Vitro Diagnostics
- Quality Assurance
Background:
- External quality assurance (EQA) is crucial for evaluating in vitro diagnostic (IVD) device analytical performance.
- Assessing device fitness for clinical use requires robust performance metrics.
Purpose of the Study:
- To develop and apply a novel method for assessing IVD device analytical performance using reclassification rates at clinical decision limits.
- To evaluate the performance of common haemoglobin A1c (HbA1c) IVD devices using this new method.
Main Methods:
- A commutable EQA program provided analytical performance data.
- Simulated 10,000 HbA1c measurements per subject using common IVD devices and the NHANES device.
- Calculated incremental reclassification rates (IRR) by comparing IVD device reclassification rates to the NHANES device, accounting for regression to the mean.
Main Results:
- Median IRRs were optimal for diabetes diagnosis (0.11–0.22%) and desirable for diabetes monitoring (-0.51–1.01%) across most tested HbA1c IVD devices.
- One device (Afinion) failed the minimal standard for prediabetes diagnosis (8.62%), while others met desirable standards (1.43–2.69%).
Conclusions:
- The incremental reclassification rate (IRR) method effectively identifies IVD device performance variations.
- EQA providers should consider reporting IRR for analytes to flag poorly performing IVD devices, especially for prediabetes screening.
Related Concept Videos
Classification of Illness
An illness is a response to a disease in which the person's level of functioning is changed compared with a previous level. The general classification of illness includes acute and chronic.
Acute illness is severe and...
Methods of Documentation VI: Case Management Model
For example, a patient with a chronic illness...
Heart Failure IV: Classification and Diagnostic Evaluation
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast, controlled...