Dual-scan conformal cone-beam CT for targeted image-quality improvement using dynamic collimation
Yuxiang Liu1, Siqi Yuan1, Xin Feng1
1National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Journal of Applied Clinical Medical Physics
|June 24, 2026
Summary
This study introduces a dual-scan conformal cone-beam CT (CBCT) method using dynamic collimation. The technique enhances image quality in target regions for radiation therapy without increasing patient dose.
Area of Science:
- Medical Imaging
- Radiation Oncology
- Image Reconstruction
Background:
- Conventional full-field Cone-Beam CT (CBCT) in radiation therapy can lead to scatter, reducing soft-tissue contrast and limiting accuracy.
- A need exists for imaging strategies that improve image quality in specific regions while retaining full-field anatomical data for patient setup.
Purpose of the Study:
- To develop and evaluate a dual-scan conformal CBCT method utilizing dynamic collimation.
- The goal is to enhance image quality within clinically relevant areas without increasing overall radiation dose.
Main Methods:
- A dual-scan protocol combining a low-dose full-field-of-view (FOV) scan with a higher-dose conformal scan was implemented.
- Dynamic collimation, shaping the X-ray field based on the target region, was simulated using multi-leaf collimators and tested physically with a kV collimator.
- Reconstruction was performed using a regularized weighted least-squares algorithm with anisotropic total-variation regularization.
Main Results:
- The dual-scan conformal CBCT method demonstrated superior image quality compared to conventional FDK reconstruction across phantom and patient data.
- Significant improvements were observed in Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR) for target regions.
- For instance, CNR increased from 0.514 to 1.463 in the FORBILD phantom and from 1.036 to 2.379 in the anthropomorphic head phantom.
Conclusions:
- The proposed dual-scan conformal CBCT strategy effectively improves local image quality in the target region.
- This method preserves full-FOV information and offers a practical approach for patient-specific CBCT guidance in adaptive radiotherapy.