Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
[Test-Retest Reliability Assessment of CT Subjective and Objective Resolution under Differentiated Scanning
1Department of Medical Equipment, Shanghai Sixth People's Hospital Affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai, 200233.
None:
Objective Based on existing CT performance testing methods, the test-retest reliability of subjective and objective results for both spatial resolution and low-contrast detectability were evaluated and analyzed under different scanning conditions. Methods A prospective research method was adopted, using clinical routine scanning protocols and parameters for the head, chest and abdomen. Eight CT scanners from different regions underwent two rounds of testing, with each round tested three times. The coefficient of variation (CV) and intraclass correlation coefficient (ICC) were used to evaluate inter-CT test-retest reliability. Bland-Altman analysis was used to evaluate intra-CT test-retest reliability. Results For head scanning, the CV range for spatial resolution was 8.85%-21.66%, with ICCs from 0.26 to 0.72; the CV range for density resolution was 5.53%-47.84%, with ICCs from 0.16 to 0.75. For chest scanning, the CV range for spatial resolution was 9.13%-19.65%, with ICCs from 0.41 to 0.75; the CV range for density resolution was 12.25%-38.89%, with ICCs from 0.18 to 0.68. For abdomen scanning, the CV range for density resolution was 5.91%-37.33%, with ICCs from 0.23 to 0.77. For the same CT, the successive results of subjective spatial resolution from head protocols demonstrated statistical significance under the first enhancement strength. Moreover, the successive results of objective spatial resolution from chest protocols demonstrated statistical significance under the first, second and third enhancement level. No statistical significance was observed under other scanning conditions. Conclusion The objective contrast evaluation method demonstrated better variability and consistency compared to the subjective evaluation method under most test conditions. Even for the same CT with iterative reconstruction algorithm at the same levels, there may exist statistical significance of the spatial resolution values from successive head and chest scanning. Cautions should be taken when test results of objective spatial resolution are analyzed.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT