Percutaneous Extraction of a Long-Standing, Inadvertently Malpositioned Left-Atrial Lead during CIED Infection
Tianbao Yao1, Peng Nie1, Zhenghao Wang1
1Division of Cardiology, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China.
Transvenous lead extraction is safe and feasible for pacemaker leads in the left heart for extended periods. This case report highlights successful removal of a 17-year-old malpositioned lead during device infection.
Area of Science:
- Cardiology
- Medical Devices
- Infectious Diseases
Background:
- Transvenous lead extraction (TLE) is crucial for managing cardiac implantable electronic device (CIED) complications.
- Limited data exists on TLE of leads with long-term malposition in the left heart, especially during infection.