Evaluation of the accuracy of pulse oximeters in real clinical settings
Theera Leeudomwong1, Tassanai Sanponpute2
1Department of Industrial Physics and Medical Instrumentation, King Mongkut's University of Technology North Bangkok, Bangkok, Thailand.
Abstract:
Introduction.Pulse oximeters are indispensable devices for noninvasive monitoring of arterial blood oxygen saturation (SpO2); however, verification of their accuracy still relies on invasive CO-oximetry performed under controlled desaturation protocols, which are impractical for routine clinical use. Although functional testing using pulse oximeter simulators is widely employed as a surrogate approach, such simulators generally lack direct metrological traceability to CO-oximetry.Objectives.This study evaluated the accuracy of five commercial pulse oximeters under routine clinical conditions at three hospitals, using site-specific CO-oximeters as reference analyzers. The objectives were to characterize real-world pulse oximeter accuracy without induced desaturation, to examine the potential for metrological traceability of arterial blood oxygen saturation measurements obtained from pulse oximeters with respect to the gold standard, and to assess the suitability of selected pulse oximeters as candidate secondary reference devices for traceable and noninvasive verification.Approach.A total of 488 paired SpO2-SaO2measurements were collected from 148 hospitalized patients during routine care. Accuracy metrics, including bias, mean absolute error, root-mean-square accuracy (Arms), Bland-Altman limits of agreement, and concordance correlation coefficient, were calculated for each site and for a pooled dataset.Results.Across all reference systems, the five pulse oximeters exhibitedArmsvalues between 2.1% and 2.7%, well within the acceptance limits specified by ISO 80601-2-61 (4%) and the U.S. Food and Drug Administration (3%). Measured accuracy remained broadly consistent across pulse oximeter models and reference CO-oximeters when standardized handling procedures were applied. Only minor inter-site differences were observed, demonstrating that consistent multicenter evaluation is feasible under routine clinical conditions within the SaO2range of approximately 80%-100%. These findings provide preliminary support for the future evaluation of selected pulse oximeters as candidate secondary reference devices and for the development of traceable, noninvasive verification approaches using calibrated simulators.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: