Related Experiment Video
Updated: Jun 27, 2026

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Performance of wrist-worn home sleep apnea testing (watch-PAT) among individuals with chronic insomnia: a comparative
Chuan Shi1, Jinmei Luo1, Rong Huang1
1Department of Pulmonary and Critical Care Medicine, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, Beijing, China.
Background:
While comorbid insomnia and sleep apnea are frequently encountered in clinical practice, the accuracy of home sleep apnea testing in this particular scenario has long been overlooked. Our study aims to evaluate the performance of a wrist-worn device in diagnosing sleep apnea among individuals with chronic insomnia.
Methods:
Participants with chronic insomnia suspected of having sleep apnea were consecutively enrolled at a sleep center. Single-night Watch-PAT 200 monitoring was conducted in parallel with attended in-lab polysomnography.
Results:
Recordings from 44 participants (41% female, age 49.1 ± 12.7 years) with a wide range of sleep duration (222.9 ~ 461.4 min) and severity of sleep apnea [apnea-hypopnea index (AHI) 0.3 ~ 83.9/h] were analyzed. Using AHIPAT ≥ 15/h as the threshold, the sensitivity, specificity, and area under the receiver operating characteristic curve for identifying moderate-to-severe obstructive sleep apnea (OSA) were 77.3% (95% CI 54.2-91.3%), 100.0% (95% CI 81.5-100.0%), and 0.92 (95% CI 0.80-0.98), respectively. Diagnostic agreement defined by clinically oriented criteria was reached in 36 (82%) subjects. The Watch-PAT overestimated total sleep time and rapid eye movement sleep duration by 19.9 min (limits of agreement -48.1 ~ 87.9 min) and 37.2 min (limits of agreement -26.1 ~ 100.6 min), respectively.
Conclusion:
Among individuals with chronic insomnia, the wrist-worn home sleep apnea testing maintained acceptable diagnostic performance for identifying moderate-to-severe OSA. Sleep duration and staging parameters provided by the device should be interpreted with caution.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
08:36Collecting Sleep, Circadian, Fatigue, and Performance Data in Complex Operational Environments
Published on: August 8, 2019
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Sleep-Wake Cycles
NREM Sleep
NREM sleep comprises four progressive stages that seamlessly merge:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Insomnia
Multiple factors contribute...
Management of Insomnia