Related Experiment Video
Updated: Jun 29, 2026

Home-Based Monitor for Gait and Activity Analysis
Published on: August 8, 2019
Validating an Activity Monitor for Acute Care Patients: The Patient/Resident Mobility Tracker (PREEMPT).
Ann Tuzson1, Christopher M Moore1, Merrick Furman2
1Department of Internal Medicine, Division of Hospital Medicine, University of Virginia, Charlottesville, Virginia, USA.
The Patient Resident Mobility Tracker (PREEMPT) accurately measures physical activity in hospitalized patients and healthy older adults. This validated device is comfortable and well-received, showing high usability for mobility monitoring.
Area of Science:
- Gerontology
- Biomedical Engineering
- Rehabilitation Science
Background:
- Accurate measurement of physical activity is crucial for hospitalized patients and older adults.
- Existing activity monitors may not be suitable for the unique needs of inpatient settings.
- The Patient Resident Mobility Tracker (PREEMPT) was developed to address these limitations.
Purpose of the Study:
- To validate the accuracy and usability of the PREEMPT activity monitor.
- To assess the device's suitability for hospitalized individuals and healthy older adults.
- To compare PREEMPT-derived mobility metrics with objective video data.
Main Methods:
- Ten healthy older adults and 46 hospitalized patients participated in the study.
- Participants wore the PREEMPT device during various physical activities.
- PREEMPT data for stride length, steps, stepping time, sedentary time, and activity time were compared to video recordings.
- Participant feedback on device usability was collected using Likert scale questions.
Main Results:
- PREEMPT measurements showed statistical equivalence to criterion data across all five mobility metrics (geometric mean ratio 0.97-0.99).
- The device demonstrated high accuracy with statistically significant p-values ranging from 0.043 to <0.001.
- The PREEMPT device achieved an average usability score of 9.4/10, indicating comfort and positive reception.
Conclusions:
- The prototype PREEMPT device demonstrates initial validity and accuracy in both healthy older adults and hospitalized patients.
- The findings support the PREEMPT device's capability to accurately measure physical activity in a relatively immobile inpatient population.
- The device is a promising tool for monitoring mobility in clinical and research settings.
More Related Videos
04:13Using a Real-Time Locating System to Measure Walking Activity Associated with Wandering Behaviors Among Institutionalized Older Adults
Published on: February 8, 2019
06:49Evaluation of a Smartphone-based Human Activity Recognition System in a Daily Living Environment
Published on: December 11, 2015
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Holter Monitor: 24-Hour Monitoring