Deep learning reconstruction dual-energy computed tomography for gastrointestinal system tumors: low-kiloelectron
1Department of Radiology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou 450052, China; Henan International Joint Laboratory of Medical Imaging, Zhengzhou, China; Henan Key Laboratory of Image Diagnosis and Treatment for Digestive System Tumor, Zhengzhou, China; Henan Key Laboratory of CT Imaging, Zhengzhou, China.
Aim:
To qualitatively and quantitatively compare dual-energy computed tomography (DECT)-derived 55 keV virtual monochromatic images (VMIs) using deep learning image reconstruction (DLIR) with 70 keV VMIs using adaptive statistical iterative reconstruction (ASiR-V) in the Gastrointestinal (GI) system tumors and to investigate whether DLIR can influence the measured iodine parameters.
Materials And Methods:
This prospective, observational study included 47 patients with GI system tumors who underwent DECT. Qualitative and quantitative assessment was performed on 70 keV VMIs using ASiR-V40%, and 55 keV VMIs using DLIR at low (DLIR-L), medium (DLIR-M), and high strength (DLIR-H). Quantitative evaluation included calculation of the contrast-to-noise ratio (CNR) and the signal-to-noise ratio (SNR). Overall image quality, lesion conspicuity, and diagnostic confidence were evaluated qualitatively by 2 radiologists. Iodine parameters of tumor lesions were compared among four groups of algorithms.
Results:
There were no significant differences in the image noise between DLIR-M and ASiR-V40%, whereas image noise was significantly lower in DLIR-H than ASiR-V40% and DLIR-M (both phases, P < 0.001). DLIR-M and DLIR-H had significantly higher CNR and SNR compared with ASiR-V40% on both phases (all P ≤ 0.001). When compared to ASiR-V40%, DLIR-M and DLIR-H had similar or improved qualitative assessment metrics. DLIR-L yielded the highest image noise and the lowest overall image quality score among these algorithms. Lastly, the measured iodine parameters were equivalent among different algorithms (all P > 0.1).
Conclusion:
Abdominal DECT at 55 keV with DLIR (-M and -H) provided qualitative and quantitative performance for GI tumor evaluation comparable to or exceeding that of ASiR-V 40% at 70 keV.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT