Updated: Jul 1, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
HIP COMPASS®: A Mechanical Intraoperative Navigation Guide Associated with Improved Revision-Free Implant
Yoshiaki Tei1, Norio Imai2, Keishi Kimura1
1Division of Orthopedic Surgery, Department of Regenerative and Transplant Medicine, Niigata University Graduate School of Medical and Dental Sciences.
Abstract:
HIP COMPASS® has demonstrated improved accuracy in intraoperative support for the acetabular component insertion angle. We assessed long-term revision-free implant survivorship following total hip arthroplasty (THA) using HIP COMPASS, with ≥10 years of follow-up. We retrospectively analyzed 210 hips of 186 patients who underwent cementless THA with ceramic-on-ceramic bearing couples, with or without HIP COMPASS. The mean follow-up duration was 10.8 years in the HIp COMPASS group and 12.7 years in the control group. Overall, 86.0% and 78.2% of cups were placed within Lewinnek's safe zone with and without HIP COMPASS, respectively. Variability in radiographic inclination and anteversion was greater in the control group than in the HIP COMPASS group. Dislocation rates were 3.2% in the control group and 1.2% in the HIP COMPASS group, with no significant difference between groups. The control group had a significantly higher revision rate than the HIP COMPASS group (7.3% vs. 1.2%). Use of HIP COMPASS was associated with improved long-term revision-free implant survivorship over ≥ 10 years and more consistent acetabular component positioning at an appropriate insertion angle, which may contribute to stable initial fixation.