Ultra-low contrast agent dosage in photon-counting CT angiography of the thoracoabdominal aorta
Jan-Lucca Hennes1, Kristina Krompaß1, Henner Huflage1
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany.
Purpose:
To assess the viability of an ultra-low contrast agent protocol (ULC) for aortic photon-counting CT angiography (PCD-CTA).
Materials And Methods:
For this retrospective, single-center study, 31 consecutive patients receiving PCD-CTA with 48 ml Imeron® 150mgI/ml (≙ 7.2 g iodine) were assessed. Virtual monoenergetic images with 45/55 keV were compared to a control sample scanned with radiation-dose-equivalent 55 keV protocols (120kVp; image-quality-level 42) employing standard (SC; 60 ml Imeron® 350mgI/ml ≙ 21 g iodine) and low contrast agent dose (LC; 40 ml Imeron® 350 mg/ml ≙ 14 g iodine). Luminal attenuation and image noise were measured, while contrast-to-noise ratios were calculated. Five radiologists rated overall image quality, luminal attenuation, and diagnostic confidence using a 5-point rating scale.
Results:
Radiation dose and body-mass-index did not differ significantly among groups (p > .999, each). ULC scans displayed significantly lower CNR (55 keV: 12.2 ± 2.7; 45 keV: 20.0 ± 4.6) than SC (35.3 ± 7.5) and LC studies (27.4 ± 10.7; each p ≤ .001) due to reduced luminal attenuation (ULC-55keV: 211.9 ± 27.3HU; ULC-45keV: 302.9 ± 42.8HU; SC 504.3 ± 70.9HU; LC: 380.8 ± 98.3HU; p < .001). Overall image quality of protocols was rated similar (each 4 [4-4]; p ≥ .395), while luminal attenuation of ULC-45keV (4 [4-4]) was rated inferior to SC (5 [5-5]; p < 0.001) but not to LC (4 [4-5]; p = 0.25). Diagnostic confidence was good for ULC (45 keV: 4 [4-4]); 55 keV: 3 [3-4]) and excellent for SC (5 [5-5]) and LC (5 [4-5]).
Conclusion:
Despite inferior contrast, aortic PCD-CTA with ultra-low contrast agent dosage at 45 keV allows for diagnostic image quality. The proposed scan protocol may hold potential for kidney protection in vulnerable patient groups.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...