Syncope due to Pacing Inhibition During VT Detection in a Conduction System Pacing ICD Programmed in AAI(R) Mode.
Rubén García1, Álvaro Riesgo1, Raquel Tornero1
1Arrhythmia Unit, Donostia University Hospital, Donostia, Spain.
Unconventional pacemaker configurations, like left bundle branch area pacing with an implantable cardioverter-defibrillator (LBBAP-ICD), may interact with device algorithms. This can cause dangerous pauses in patients dependent on pacing.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Dedicated generators for conduction system pacing are limited.
- Unconventional device configurations are sometimes necessary for specific patient needs.
- Conduction system pacing offers potential advantages over traditional pacing.